20 Most Recent Invacare Supply Group Invacare Pronto M51 Power Wheelchair - Blue - Power wheelchair Questions & Answers

0helpful
3answers

How do I charge My Pronto M51 motorized wheelchair?

Margie,   Yes you may charge it overnight. If you use it everyday,charge it every night. Once the charger senses 24 volts it stops sending a charge,so they cannot be overcharged.  I suggest when you go to bed put on the charger and unplug it when you get up in the morning.
10/2/2020 2:12:36 AM • Invacare Supply... • Answered on Oct 02, 2020
0helpful
3answers

My tool light is blinking on my electric

How many times is it blinking?
5/12/2014 11:58:12 PM • Invacare Supply... • Answered on May 12, 2014
0helpful
1answer

Someone gave me a pronto m51 sure step motorized

Jo, nope you are safe. Most of the chargers are designed to cut off and go into trickle charge once they sense that the batteries are fully charged. The problem with overcharging comes after 18 hours or so. You should never leave the charger plugged into the chair for more than 24 hours. On the M51 the charger is built into the chair so just plug it in when you go to bed and unplug it when you get up and you'll be ok. You can check for full charge by looking at the charger there should be a red light and a green light when fully charged. Hope this helps
5/5/2014 8:15:53 PM • Invacare Supply... • Answered on May 05, 2014
0helpful
1answer

Will only work at a slow speed speed con worked

have a dealer tech check it with a programer and he will tell u what the problem is by checking the error log could be just a joy stick needs to be reset or the control box in the rear,make sure your insurance will help with repair
1/20/2014 12:30:37 PM • Invacare Supply... • Answered on Jan 20, 2014
0helpful
2answers

Invacare pronto m51 moving very

Check that the speed control on your joystick is set to high. If it is then you probably have a battery problem. Even with all the lights lit on the joystick you can have one battery of the two on the chair that is not charging fully. This will cause the M51 to creep and not work correctly
12/15/2013 11:10:04 PM • Invacare Supply... • Answered on Dec 15, 2013
0helpful
2answers

Tool; light blinks 9 times battery is charged

the batteries were dead
7/3/2013 4:35:28 PM • Invacare Supply... • Answered on Jul 03, 2013
0answers
0helpful
3answers

Wheelchair problem

sounds like you might have locked joystick by mistake. Try pressing and holding down the on/off switch for 5 seconds.
12/19/2012 7:16:19 PM • Invacare Supply... • Answered on Dec 19, 2012
0helpful
1answer

M51 charger has solid red light and a blinking green light

Battery is shorted or may have cristlized replace battery. This will solve your problems.
9/16/2012 11:16:04 PM • Invacare Supply... • Answered on Sep 16, 2012
0helpful
1answer

Footplate on M51 keeps falling down.

Sounds like you need to tighten the bolt that attaches the footplate to the bracket. You will need a 5/16" allen wrench for the bolt and a 14mm socket for the lock-nut. I'm surprised the footplate lasted 2 years without loosening up. Typically I have to tighten that bolt for customers once every few months. Hope this helps.
10/29/2011 2:34:05 PM • Invacare Supply... • Answered on Oct 29, 2011
0helpful
1answer

TDX SP; the screen reported

Hi,
Seems the Control Panel is not connected properly on the unit or may have gone lose check the control panel cable connection on the unit.
Make sure that when check the connection it is turned OFF and not plugged in or charging to avoid any short or shock.
See below for info on how to check the control panel connection,
mvincentc_52.jpg

See picture below to check the controller connection form the battery.
If the issues till exist then probably the controller program is corrupted or may need reprogramming on the unit.
The reprogramming can only be done by a service technician of Invacare for the reprogramming of the controller if you still have the Manual supplied with the wheel chair it should have the info ther and number to call, if ever the connection is not the issue.
I hope this helps.
Have a nice day,
mvincentc
8/4/2011 1:15:06 AM • Invacare Supply... • Answered on Aug 04, 2011
0helpful
1answer

PRONTO M51 FEELS LIKE I AM DRIVING OVER ROUGH ROAD

If you chair was down for a long time before getting new batteries the tires could have developed flat spots on them. Also could have brush problems in the motors that make the motors run jerky from sitting too long.
5/1/2011 12:05:19 AM • Invacare Supply... • Answered on May 01, 2011
0helpful
1answer

Will only work at a slow speed, speed control has

The joystick sounds like its off alignment,Once knocked off alignment it will have to be replaced .

hope i was able to help
1/21/2011 4:00:22 PM • Invacare Supply... • Answered on Jan 21, 2011
0helpful
1answer

Do not know if chair is taking a charge lights

hi let me see if i can help ,

When the lights go back and forth on your joystick that means the chair knows its pluged in .Now the chair itself has a built in charger that is mounted on the back of your chair .The back of the chair ,there is a opening a cutout in the frame .when you plug in your chair there will be a red light and a yellow ,yellow means the chair is charging, wait till it turns green,this means fully charged.When you plug in the chair youll see the lights through that hole.Look for the lights.

I recommend my customers to leave there chair charging over night look at the light it will be green when you turn the chair on the joystick should show fully charge.


Good luck hope i was of some help.
1/20/2011 2:10:54 AM • Invacare Supply... • Answered on Jan 20, 2011
0helpful
1answer

My prontom51 willnot move trouble light

The Invacare Pronto M51 with a flash code of 6 indicates a right brake fault. This can mean several things and there are several solutions to this problem.
The brake lever can be in the wrong position and the brake can be in the release position. Work the brake lever up and down and see if this clears the fault and the chair drives.
There could be a bad connection from the motor to the controller in the base. Check the wiring for cuts and breaks. If you can get to the wiring plugs from the motor (there are two) check the connections and plugs for corrosion and a good connection.
To verify that you have a bad brake you need a digital voltmeter to check the resistance of the electromagnetic brake coil. It should be about 40 to 60 ohms across the two white connectors. You can check the motor at the same plug. If you have made these checks and the motor and brake test good then you have a bad controller output to that motor. This can be tested by swapping the motor outputs from the controller and see if the error code goes to the left motor output. The controllers run about $1000 retail and a technician can verify this a second time by taking a good known controller and plugging it in to see if the chair now runs.
In my experience if the motor/brake/gearbox assembly is more than 3 years old have it replaced. That being said, an experienced technician should verify a bad brake/motor/gearbox assembly first. On this chair, the only cost effective repair is to replace the entire brake/motor/gearbox assembly. The retail price is about $700. Keep in mind that if the right assembly is worn out the left is not far behind. Typically if insurance will pay for the repair, I opt to replace the left and right brake/motor/gearbox assemblies at the same time so you do not pay for the labor and inconvenience twice.
Try the simple fixes mentioned first, and then have a technician evaluate the chair for the other problems I have discussed. Hope this helps.
12/8/2010 3:52:40 AM • Invacare Supply... • Answered on Dec 08, 2010
Not finding what you are looking for?
Invacare Supply Group Invacare Pronto M51 Power Wheelchair - Blue - Power wheelchair Logo

44 questions posted

Ask a Question

Usually answered in minutes!

Top Invacare Supply Group Health & Beauty Experts

Rob Hill
Rob Hill

Level 3 Expert

1480 Answers

john h

Level 3 Expert

29494 Answers

Steve Pack
Steve Pack

Level 3 Expert

7126 Answers

Are you an Invacare Supply Group Health and Beauty Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

View Most Popular

Invacare Supply Group Health & Beauty

  • Invacare Supply Group Health & Beauty

Most Popular Question

tool light blinking electric

  • Health & Beauty
Loading...