Ecg1200, 10mins delay when switched it on before it starting on. the battery looks fine.
Shenzhen Biocare Electronics Co., Ltd, ecg machine 1200 model. WHEN SWITCHED ON the display shows biocare but after that it delays for 5-10mins before the wave patterns appears.
- If you need clarification, ask it in the comment box above.
- Better answers use proper spelling and grammar.
- Provide details, support with references or personal experience.
Tell us some more! Your answer needs to include more details to help people.You can't post answers that contain an email address.Please enter a valid email address.The email address entered is already associated to an account.Login to postPlease use English characters only.
Tip: The max point reward for answering a question is 15.
hi, if you can provide your email id then i will sent you the service manual for Mac 1200 ecg. Its not possible to upload the service manual here . Its in pdf format. Ajin
hi,
There is a 3V batterty inside the machine located in the main board. If the battery is faulty then the memory will not hold. You can replace the battrey by removing the back cover. Have adde d a picture. The part number will be witten on it. Take the battery to any nearest electronic sparepart shop and by a new one . When you insert the new battery remember one thing the the possitive side of the battery should be on top.
ok
hi,
You have to change the battery of you ecg machine. Because even if the power is connected the ecg machine works on battery. If the battery is low the machine will go OFF after booting.
ok
Is the paper correct for the machine (look at old ECG's & check paper sensor black mark is in same place) also check paper is not in wrong way round or upside down?
Bob - www.woodenmen.co.uk
×